Manual number:
ITKU-09-03-11-11-A
PUE C41H Indicator
for cooperation with load cells
MANUFACTURER OF ELECTRONIC
WEIGHING INSTRUMENTS
RADWAG 26 – 600 Bracka 28 Street - POLAND
Radom, phone +48 48 384 88 00, phone/fax +48 48 385 00 10,
Sales Department +4848 366 80 06
www.radwag.com
Zobrazit stránku 0
1 2 3 4 5 6 ... 162 163

Shrnutí obsahu

Strany 1 - PUE C41H Indicator

Manual number: ITKU-09-03-11-11-A PUE C41H Indicator for cooperation with load cells MANUFACTURER OF ELECTRONIC WEIGHING

Strany 2

4. MAIN DIMENSIONS Main dimensions of PUE C41H 5. DESCRIPTON OF CONNECTORS Terminal connectors 1 – I/O connectors 2 – RS232, RS485 connecto

Strany 3 - Table OF CONTENTS

• Put a load to be a standard mass on the weight pan, • After the equilibrium is reached () press , • You will see the indication of 100,00%,

Strany 4

From this moment all results will be displayed in percents. Return to weighing: See 11.3. 24.7. Weighing animals 24.7.1. Weighing time settin

Strany 5

24.7.2. Starting the work mode • WGH. ANIMALS mode is signalled by showing ► pictogram in the right side of the picture. 24.7.3. Procedure o

Strany 6

24.8. Dosing Standard indicators are equipped with signalling fields and internal circuit of - 3 optoinsulated inputs (5÷24V DC), 3 optoinsulated r

Strany 7 - 1. INTENDED USE

24.8.2. Time interval between changing dosage thresholds By setting <P7.4.2 DELAY> parameter users can set a delay time between changing dosi

Strany 8 - 2. PRECAUTIONARY MEASURES

Using numeric keys inscribe a requires value (0 to 60 s) and confirm by pressing . Return to weighing: See 11.3. 24.8.4. Mode for OUTPUTS Soft

Strany 9 - 3. WARRANTY CONDITIONS

Caution: If dosing is initiated when the indication is > LO , after reaching the indication below LO output (outputs)will be automatically switch

Strany 10 - 5. DESCRIPTON OF CONNECTORS

24.8.6.1. Correction mode Procedure: • Enter <P7 WORK MODES> and then: NONE - Correction function disabled CONSTANT - Dosing refered to a

Strany 11 - 7. GETTING STARTED

• Use the scale keyboard a starting correction value and confirm it by pressing . Notice: 1. The starting correction value is estimated experim

Strany 12 - 9. PICTOGRAMS

• Use a keyboard to enter a maximum value for the process and confirm it by pressing . Notice: The introduced maximum value for corrections is inte

Strany 13 - 10. FUCTIONS OF KEYS

Notice: In accordance to the number of mounted modules the number and the placement of glands and connectors can vary. Connectors and glands mentione

Strany 14 - 11. MENU - PARAMETERS

24.8.7. Starting work modes After entering the DOSING mode the pictogram is displayed. Press to start a dosing process. After starting it s

Strany 15

Pressing will cause the process to continue. Pressing will cause the process to break. Return to weighing: See 11.3. 25. DIAGRAMS OF CONNECTIO

Strany 16

Scale – barcode scanner (LS2208) Scale – printer (CITIZEN, EPSON) cable Scale – Zebra printer cable - 112 -

Strany 17

Scale - 3IN/3OUT (RSTS-8-184/2M) cable PRINT, TARA, ZERO external buttons cable 26. CONNECTORS Caution: In accordance to the number of moun

Strany 18

INPUTS OUTPUTS description wire colour description wire colour GND IN RED COMMON YELLOW IN1 BLUE OUT1 GREEN IN 2 PINK OUT 2 BROWN IN 3 GRE

Strany 19

27. SPECIFICATION OF ADDITIONAL MODULES Apart from standard interface, it is possible to equip terminals with additional module increasing functiona

Strany 20 - 12. WEIGHING

Main board view with connectors for additional modules: J18 – DP 1 module J1 – Ethernet ET 1G, ET 1 D module J3 – Relay module J12 – WE 8

Strany 21

- 117 - This module is designed according to TCP/IP 10/100 Mbit/s standard. It comprises two signalling LED’s: • D2 lights - Ethernet connection es

Strany 22

9. Connect the PT0014 cable to J2 and J3 connectors on the Ethernet module according to the description; 10. Cable (twisted pair) or wires from the E

Strany 23

27.1.2. Drawings of sockets and cables for Ethernet Ethernet socket for ET 1G version Ethernet cable for ET 1D (version for a SWITCH) 27.2.

Strany 24

8. KEYBOARD 9. PICTOGRAMS No Pictogram Description 1. Zero indication (Autozero zone) 2. Equilibrium 3. kg (g) Weighing mode 4. Battery

Strany 25 - 13. MAIN PARAMETERS

Module accessible in three configurations: • Voltage output AN 0-10V • Current output AN 4-20mA • Current output AN 0-20mA 27.2.1. Technical

Strany 26

Mounting of AN module on the main board of PUE C41H 27.2.3. Configuration of work modes of analogue modules A work mode of analogue modules can

Strany 27

27.2.4. Connections to AN module Drawing of connections of voltage output: COLOURS OF WIRES Colour Signal Pink UOUT + Gray GND Yellow +24V DC Gree

Strany 28

27.3. Relay module - PK1 Relay module PCB - PK1 This is an alternative solution for reed relay outputs present on the main board in the standard

Strany 29

Installing procedure: 1. Unplug the scale from mains; 2. Unscrew and take off the back wall of the housing; 3. Install your module in J3 on the m

Strany 30

27.3.3. Drawing of cables and outputs Relay outputs diagram: SIGNALS AND DESIGNATIONS OF CONDUCTORS Wire number Description 1 OUT 1 2 OUT 2 3 OUT

Strany 31 - 14. PORTS PARAMETERS

- 126 - 27.4.2. Colours of cables for I/O INPUTS OUTPUTS wire number description wire number description 1 IN 1 6 OUT 1 2 IN 2 7 OUT 2 3 IN

Strany 32

Installing WE4 modules on the main board of PUE C41H 27.5. WE 8 - 8 inputs / 8 outputs module 8 inputs / 8 outputs PCB - WE 8 WE 8 modules c

Strany 33

- 128 - 27.5.1. Technical specification Parameters of outputs Quantity of outputs 8 Type of outputs Reed operation contacts Wire diameter 0,14 -

Strany 34

A WE 8 module placement on the main board of PUE C41H 27.5.3. I/O diagram WE8 inputs diagram WE8 outputs diagram - 129 -

Strany 35

9.1. Battery charge indication pictogram is situated in the upper right corner informed about the discharge level or charging process: • pictogr

Strany 36 - 15. EXTERNAL DEVICES

27.5.4. Description of input output wires Signals led out with two cables10x0,5mm2 with numbered conductors. INPUTS OUTPUTS Wire number descriptio

Strany 37

- 131 - 27.6.1. Technical specification Useful number of internal divisions 8 388 608 OIML class III Number of verification divisions 6 000 Maxim

Strany 38

27.6.3. Connecting additional platforms Connecting 6-wire tensometers E - tensometer cable shield REF+ - “SENSE +” from tensometer (JP1 no

Strany 39

Connecting 4-wire tensometers E - tensometer cable shield REF+ - solder jumper JP1 REF- - solder jumper JP2 IN+ - “OUTPUT+” from tenso

Strany 40

platforms connected to indicators in metal housing via a cable only platforms electrically connected to indicators’ metal housings e.g. pillars, ra

Strany 41

4. During installation turn your attention to plastic columns. They should be placed one side in mounting holes in the main board and the other side

Strany 42

27.7. RS485 led out via RS 1D gland A version with the RS485 interface led out via a gland (in the standard solution RS485 is present in a sock

Strany 43

PT0012 cable connecting to the main board of PUE C41H 5. Connect the PT0012 shield to the housing (4mm screwed terminal on the back wall) 6. Th

Strany 44

- 138 - 28. COMMUNICATION PROTOCOL 28.1. General information A. A character protocol scale-terminal has been designed for communication between R

Strany 45

- 139 - UH Set upper threshold ODH Read lower threshold OUH Read upper threshold PC Send all implemented commands Notice: 1. Each command have to

Strany 46

Selection purpose variables … N/A Work mode selection N/A Log out Inscribing MIN, MAX thresholds View of statistics … N/A … Programmable k

Strany 47

- 140 - Z_A CR LF Z_^ CR LF - command accepted and in progress - command comprehended but zero range overflow appeared Z_A CR LF Z_E CR LF - command

Strany 48 - 16. DATE / TIME SETTING

- 141 - 28.4.4. Set tare value Syntax: UT_TARE CR LF, where TARE – tare value Possible answers: UT_OK CR LF - command executed UT_I CR LF - comma

Strany 49

- 142 - 28.4.6. Send the result immediately in basic unit Syntax: SI CR LF Possible answers: SI_I CR LF - command comprehended but cannot be exec

Strany 50 - 17. PRINTOUTS

- 143 - n - weighing platform number mass - 9 characters justified to the right unit - 3 characters justified to the left Example: Let us assu

Strany 51 - LAST STB

- 144 - Possible answers: SUI_I CR LF - command comprehended but cannot be executed SUI_A CR LF MASS FRAME - command accepted and in progress - mas

Strany 52

- 145 - 28.4.12. Switch on continuous transmission in current unit Syntax: CU1 CR LF Possible answers: CU1_I CR LF - command comprehended but cann

Strany 53

- 146 - 28.4.15. Unlock the scale keyboard Syntax: K0 CR LF Possible answers: K0_OK CR LF – command in progress 28.4.16. Initiating of dosing

Strany 54

- 147 - Possible answers: UH_OK CR LF - command executed ES CR LF - command not comprehended (wrong mass format) 28.4.20. Read lower threshold

Strany 55

Possible answers: PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,PC,K1,K0,DH,UH, ODH,OUH,S1,S0,OT,UT" – command executed, the indicator have sent all th

Strany 56

- 149 - Example 2: ? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - the printout generated from the scale after pressing ENTER/PRINT. Example 3: ^ _ _ _ _

Strany 57

- 15 - P 1.2.1 FITER | AVERAGE P 1.2.2 MED. FILTER | YES P 1.2.3 LO THRESH. | 20 d P 1.2.4 TARE MODE | STDRD P 1.2.5

Strany 58 - 18. DATABASES

- 150 - 29. ERROR MESSAGES ERROR ”XXX” ESC – RETURN TO PREVIOUS SETTING (where: XXX – parameter name) – confirmed with ENTER of a wrong value in the

Strany 59

31. TECHNICAL PARAMETERS Technical data: PUE C41H Display LCD Casing stainless steel Keyboard membrane Power supply 85-265V AC 50/60Hz, battery 6V

Strany 60

- 152 - 32. APPENDIX 32.1. Communication with barcode scanners 1. For communication with barcode scanners RADWAG scales use RS232 interfaces and

Strany 61

32.2. Cooperation with „EDYTOR WPW” – PC software tool Standard WPW scales can be supervised by the accompanied program EDYTOR WPW, that can work a

Strany 62

32.2.2. Setting RS232 parameters On order to establish the connection with EDYTOR WPW through RS232 interface follow the remarks below: • Run EDY

Strany 63

32.2.3. Setting Ethernet parameters On order to establish the connection with EDYTOR WPW through Ethernet interface follow the remarks below:

Strany 64

• In the overlap <Transmission setting>: − Set the communication type „Ethernet Transmission TCP”, − „IP address 1” the same as <P2.4.2

Strany 65

HOLD B. NO CHUTE ENABLE NO • Enter the submenu <P8.2 OUTPUT CONF.> and configure outputs according the table below: MIN NO OK NO MAX NO S

Strany 66

- 158 - Caution: The way how outputs operates in thresholds is set in <P7.4.4 OUTPUT MOD.>(see ch. 24.8.4). For this example this parameter is

Strany 67

Procedure of designing the report pattern: 1. Run program EDYTOR WPW and initiate communication with the scale according to ch. 32.2 of this manual,

Strany 68

- 16 - P 3.5 ADDITIONAL DISPLAY P 3.5.1 DISPL. PORT | NO P 3.5.2 DISPL. TYPE | LCD P 4 DATE / TIME P 4.1 DISPL. TIME | *

Strany 69

4. Enter the new report pattern: RADWAG WAGI ELEKTRONICZNE --------------------------------- Product: %129 Weighings: %130%089. %073 %075 %076 %07

Strany 70

32.6. Examples of surge protections • Connection diagram of dosing indicators outputs with protection for DC: - 161 -

Strany 71

• Connection diagram of dosing indicators outputs with protection for AC: - 162 -

Strany 72 - from any work

MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS PRODUCENT WAG ELEKTRONICZNYCH ,,RADWAG” 26 – 600 Radom, Bracka 28 Street POLAND Phone

Strany 73

- 17 - P 7.4.6.2 START CORRECTIONS | 0.000 P 7.4.6.3 MAX CORRECTIONS | 0.000 P 7.4.6.4 AVERAGING VALUE | 3 P 8 I/O

Strany 74 - 19. REPORTS FROM WEIGHINGS

11.2. Navigating within the menu level Use keyboard to browse the menu. 11.2.1. Keyboard Entering the main menu, special characters in the editin

Strany 75

11.2.2. Quick access It is possible to move quickly within the parameters’ menu using to . Procedure: 11.3. Return to weighing Press , un

Strany 76

- 2 - NOVEMBER 2011

Strany 77

12. WEIGHING Put a load you want to weigh on the weighing pan. When the pictogram appears it means that the result is stable and ready to read.

Strany 78 - 21. STATISTICS

• Avoid side loads, particularly side shocks should be avoided: 12.2. Tarring In order to determine the net mass put the packaging on the pan.

Strany 79

• Press , • In the lower line you will see an editing field: • Inscribe the tare value: • Press , • The scale return to weighings mode

Strany 80

12.5. Weighings in two ranges Switching between the I range and the II range happens automatically (exceeding Max of the I range). Weighings in the

Strany 81 - 22. OTHER PARAMETERS

Selection: • When the main unit is [kg], users can select among: [kg, lb, oz, ct, N, g] , for verified scales [lb, oz, N] are not accessible; •

Strany 82

13. MAIN PARAMETERS Users can adjust the scale to external ambient conditions (filtering level) or particular needs (autozero operation, tare m

Strany 83

13.2. Median filter This filter eliminates short mechanical shocks. Procedure: • Enter <P1 SCALE PARAMETERS> and then: NO - filter disabl

Strany 84

As dosing is a kind of dynamic state, which results in continuous changes in measurements, the averaged number of samples in the filter have an effe

Strany 85

Procedure: • Enter <P1 SCALE PARAMETERS > according to 11.2. of the manual: Return to weighing: See 11.3. 13.4. Minimal mass parameter

Strany 86

Return to weighing: See 11.3. 13.5. Tare function This parameter allows to set appropriate parameters for tarring. Procedure: • Enter <P1

Strany 87

- 3 - Table OF CONTENTS 1. NTENDED USE ...

Strany 88 - 23. SCALE CALIBRATION

AUTO - disable automatic tare (the mode is remembered after restart); NORMAL - tarring by pressing ; MEMORY - tare memory mode - the last tare

Strany 89

- 31 - NO - Autozero disabled YES - Autozero enabled Return to weighing: See 11.3. 14. PORTS PARAMETERS It is possible to connect external de

Strany 90

Return to weighing: See 11.3. 14.1.2. Baud rate of RS 485 Procedure: • Enter < P2 COM PORTS PARAMETERS > according to 11.2. and then:

Strany 91 - 24. WORK MODES

14.1.3. RS 232 parameters Procedure: • Enter <P2.2 RS232 (1)> and press , • Using scroll to <P2.2.2 DATA BITS> and press : • The

Strany 92 - Name Description

14.1.4. Setting of RS 485 parameters Procedure: • Enter <P2.1 RS485> and press , • Using go to <P2.1.2 DATA BITS> and press : • T

Strany 93

- 35 - 14.2. ETHERNET setting ETHERNET can be configured in <P2.4 ETHERNET>. Inventory of default parameters: No NAME VALUE DESCRIPTION

Strany 94

- 36 - Return to weighing: See 11.3. 15. EXTERNAL DEVICES 15.1. Cooperation with a computer WPW scales can cooperate with computers of IBM PC cl

Strany 95

Caution: Standard scales can communicate with computers only via RS232(1) or RS485. Return to weighing: See 11.3. 15.1.2. Type of printout scal

Strany 96

Caution: The procedure of designing non-standard printouts is described in chapter 17.6 of this manual. Return to weighing: See 11.3. 15.1.3. Ad

Strany 97

Procedure: • Enter < P3.1 COMPUTER > according to 11.2. and then: Return to weighing: See 11.3. 15.1.5. Cooperation with „E2R System” S

Strany 98

- 4 - 15.3.2. etting the START parameter...43 S 1 .3.3. Set

Strany 99

Return to weighing: See 11.3. 15.1.5.2. Buffer for weighings Users can declare the quantity of performed measurements to be saved in the inter

Strany 100

Return to weighing: See 11.3. 15.1.5.3. The lock of product change Users can lock changing products by scale operators cooperating with „E2R SYS

Strany 101

Procedure: • Enter <P3.2 PRINTER> and then: Return to weighing: See 11.3. 15.3. Cooperation with a barcode scanner The scale gives poss

Strany 102

Return to weighing: See 11.3. 15.3.2. Setting the START parameter Procedure: • Enter <P3.3 BARCODE SCANNER> and then set the START param

Strany 103

Return to weighing: See 11.3. 15.3.3. Setting the LENGTH parameter Procedure: • Enter <P3.3 BARCODE SCANNER> and then set the LENGTH param

Strany 104

Caution: In parameters <P2 COM PORTS PARAMETERS> set the baud rate for the one that requires the barcode scanner (default 9600b/s). 15.4.1

Strany 105

• In parameters <P2 COM PORTS PARAMETERS> set the baud rate (default 9600b/s). • Enter the database of operators and then find and edit the

Strany 106

Return to weighing: See 11.3. 15.5.2. Selecting an additional display type Procedure: • Enter < P3.5 ADDITIONAL DISPLAY > according to 1

Strany 107

16. DATE / TIME SETTING Enter <P4 DATE / TIME> to set these parameters. 16.1. Time view Procedure: Return to weighing: See 11.3. 16.2

Strany 108

• After pressing you will see: • Enter an appropriate value and confirm it with , • You will have to enter the following variables in sequen

Strany 109

- 5 - 22.6. Beep” sound – key-press reaction...86 “ 22.7.

Strany 110

YY - MM - DD - year - month – day YY - DD - MM - year - day - month DD - MM - YY - day - month - year Return to weighing: See 11.3. 17. PRINTO

Strany 111

NO - manual printout WHEN STB - automatic printout after stabilising CONTIN. - continuous printouts LAST STB - printing the last stable result afte

Strany 112 - Scale – Zebra printer cable

Procedure: • Enter <P5 PRINTOUTS> according to 11.2. and then: Return to weighing: See 11.3. Notice: In case of verified scales <P5.2 S

Strany 113 - 26. CONNECTORS

Return to weighing: See 11.3. 17.4. Non-standard printouts Users have possibility to design non-standard printouts in <P5.4 PRINTOUT>. Pr

Strany 114

• After pressing , you will see a cursor. Software is ready to accept your data. Non-standard printout can comprise: • Constant texts, • Vari

Strany 115

- 55 - 17.6.2. Variables appearing in all modes CODE DESCRIPTION %000 Mass in a basic unit of the active platform %001 Mass in a current unit of t

Strany 116 - Ethernet module PCB

%067 Net mass (lb) in EAN 128 %068 Gross mass EAN 128 %070 Date in EAN 128 %126 Reference quantity for counting pieces %127 Difference of tare va

Strany 117

17.6.4. Variables for printouts of reports from weighings %086 Weighing status (threshold for weighing - MIN, OK or MAX) %087 Sum of weighings %08

Strany 118

2. Standard software do not include a database of contractors. That is why the printout RP03 substitutes „Contractor code” with dashes < - - -

Strany 119 - Module of analogue outputs

Procedure: • Enter <P6 DATABASES> according to 11.2. and then: Return to weighing: See 11.3. Users can log in even if this procedure is d

Strany 120

- 6 - 27.6. DP1 – module for an additional platform...130 27.6.1. ec

Strany 121

18.2. Access level 18.2.1. Access level to edition of databases Any administrator is able to set one of three levels of edition of databases: admi

Strany 122

OPERATOR -Can perform weighings, edit P1, P4, P9 parameters; cannot edit databases and erase statistics; cannot change weighing thresholds Min, Ma

Strany 123

NUM -only digits 0 to 9 ALPHANUM - alphanumeric password Return to weighing: See 11.3. 18.4. Type of codes There is possible to select a type

Strany 124

NUM -only digits 0 to 9 ALPHANUM - alphanumeric password Return to weighing: See 11.3. 18.5. Access to edition of databases Procedure: Retur

Strany 125

18.6.1. Quick code search Procedure: • While in any work mode press (select / view products in the assortment database) and then: • Inscribe

Strany 126

• Inscribe the product name you search and then press . • The program displays the record you search in the bottom line. Caution: You can inscri

Strany 127

Return to weighing: See 11.3. 18.7. User database The database of operators can hold up to 100 records in standard setting of databases. Proced

Strany 128

Fields in an operator record: NAME Operator name (max. 40 characters) CODE Operator code (max. 6 characters) PASSWORD Password for logging (max. 8

Strany 129 - 27.5.3. I/O diagram

18.8. Assortment database The assortment database can comprise up to 3000 records. Procedure: • Enter databases according to 18.5 of this manual

Strany 130

Fields in the assortment database: NAME Product name (max. 40 characters) CODE Product code (max. 7 characters) EAN Constant code that can be used

Strany 131 - 27.6.2. Colours of wires

- 7 - 1. INTENDED USE PUE C41H indicators are intended to working in high humidity and a wide temperature range -10÷40°C. The terminal is equipped w

Strany 132

18.9. Weighings database Every result sent from the scale to a printer is also saved in the database of weighings. There is possible to save up to

Strany 133

• To delete all records press . • To print the selected record press . Notice: 1. Single record can be deleted only when first in the database

Strany 134

• You can browse the database using or by inscribing a position number using to . Caution: If users confirm a tare value greater than the ma

Strany 135

• To add or modify a record (inscribing characters like in mobile phones) press , • To add or modify a record (inscribing only digits 0 to 9) pres

Strany 136

Where: < „RADWAG”> is a universal variable No 3. After entering non-standard printouts (see 17.5) we design a printout: % 8 0 3 \ C \ T D A T

Strany 137

19.2. Printouts of reports After entering (see ch. 19.1) you will see the following display: • To edit variables press , • Confirm changes by

Strany 138 - 28. COMMUNICATION PROTOCOL

• After editing all variables go to the next item <PRINT REPORT> by pressing . • Press , you will see the window: • Using select one o

Strany 139

Procedure: • To declare a number of input ascribed to the button enter <P8.1 EXTERNAL BUTTONS> and then, • Press , you will see <P8.1.1

Strany 140

Procedure: • To declare an output number connected with the function, enter <P8.2 OUTPUT CONF.> and then: • Press , you will see <P8.2.1

Strany 141

Procedure: • Enter < P6 DATABASES > according to 11.2. and then: GENERAL - global statistics, ASSORT - statistics for every product. Retu

Strany 142

2. PRECAUTIONARY MEASURES 2.1. Precautions A. Please, read carefully this user manual before and use the device according to its intended use,

Strany 143

Using users can view the current statistics: SUM – total mass of all details, AVG – every mass of all details, MIN – minimal mass, MAX - maximal

Strany 144

Caution: 1. When a user changes a work mode all statistics are automatically deleted. 2. Statistic data are common from all platforms connected

Strany 145

22.2. LED power setting According to the requirements (e.g. intensity of external light) it is possible to change the light flux from LEDs in the sc

Strany 146

RED LEDS NONSTAB. diodes start to lit after exceeding the LO threshold (see 13.3), RED LEDS STABLE diodes start to lit after exceeding the LO

Strany 147

22.4. Automatic power down Changes can be made in <P9.3 POWER SAVE>. When the POWER SAVE function is enabled the device switches off after 5

Strany 148

22.5. Backlight According to the requirements (e.g. intensity of external light) it is possible to: • Switch on/off or set the backlight operatio

Strany 149

22.5.2. Backlight - power supply from the accumulator Procedure: • Enter <P9 OTHER> according to 11.2. and then: Return to weighing: See

Strany 150 - 30. TROUBLE SHOOTING

NO - no “beep” after pressing keys YES - “beep” after pressing keys Return to weighing: See 11.3. 22.7. Keypad modes The program allows t

Strany 151 - 31. TECHNICAL PARAMETERS

2ABC - Digits come first after pressing a key ABC2 - Letters come first after pressing a key Return to weighing: See 11.3. 22.8. Software version

Strany 152 - 32. APPENDIX

23.1. Calibration procedure • Enter <P10 USER CALIB.> according to 11.2. and then: • Unload the pan, • Press . During adjusting a start ma

Strany 153

2.3. Operation in a strong electrostatic field If the device is about to operate in a strong electrostatic field (e.g. printing houses etc.) it sho

Strany 154

23.2. Start mass adjustment It is possible to adjust only a start mass, it helps to correct the start zero when the span does not change. Procedure

Strany 155

24. WORK MODES 24.1. Accessibility of work modes In <P7.1 ACCESSIBILITY> users can declare work modes that are accessible after pressing . P

Strany 156

24.2. Programmable keys In parameter group <P7.2 KEY FUNCTIONS> users can ascribe different functions to keys: , , , or to get an easies

Strany 157

- 93 - EDIT. PRINTOUT 3 Edit non-standard printout No 3 EDIT. PRINTOUT 4 Edit non-standard printout No 4 ASSORTMENT CODE Quick search of the assortme

Strany 158

CC LABELS NO * Setting the number of CC labels to be printed MULTIPLIER * Setting the recipe multiplier RECIPE CODE * Inscribing a recipe code for a

Strany 159

• You will see the following display: • Using to numerical keys inscribe a MAX threshold and confirm with , • Software returns to WEIGHING

Strany 160

Notice: Users can use this function in other work modes like counting pieces, weighing in percents etc. Only values and units can change. 24.4. Max

Strany 161

24.5.1. Enabling work modes • COUNTING PCS procedure is active when pictogram is displayed. 24.5.2. Setting standard mass by inscribing the

Strany 162

24.5.3. Setting the standard mass by declaring the quantity of a sample Procedure: • Enter COUNTING PCS. and then: While inscribing a sample qu

Strany 163 - MANUFACTURER

Notice: If a user confirms a sample with with the empty weight pan, <Err6> will be displayed. Return to weighing: See 11.3. 24.6. Devi