Radwag PUE C41H Indicator Uživatelský manuál Strana 157

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 163
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 156
HOLD B. NO
CHUTE ENABLE NO
Enter the submenu <P8.2 OUTPUT CONF.> and configure outputs
according the table below:
MIN NO
OK NO
MAX NO
STABLE NO
THRESHOLD 1 OUT1
THRESHOLD 2 OUT2
CHUTE OUT3
2. Return to weighing according to ch.11.3.
3. Connect the cable Scale - I/O (see ch. 25) and then install external
buttons START, STOP (terminal inputs) and subsequent devices
to the terminal outputs according to the drawings below:
4. Enter the DOSING operating mode according to ch. 24.8.6,
5. Press
to enter thresholds setting:
MIN – mass value over which the bulk dosing changes to the precise
dosing ,
MAX – mass value over which dosing is completed and CHUTE starts.
6. Before you start dosing unload the pen (container, silo) or tare the
scale pressing
.
7. Start the process by pressing the external START button,
pictogram starts to blink, and threshold 1, THR 1 bulk dosing,
is performed.
- 157 -