Radwag BALANCES 3Y SERIES Uživatelský manuál Strana 213

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 246
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 212
- 213 -
Format: Z CR LF
Accessible responses:
Z_A CR LF
Z_D CR LF
- command understood and in progress
- command carried out
Z_A CR LF
Z_^ CR LF
- command understood and in progress
- command understood but zeroing range exceeded
Z_A CR LF
Z_E CR LF
- command understood and in progress
-
time limit exceeded while waiting for stable measurement
result
Z_I CR LF
- command understood but not accessible at this moment
Tare the balance
Format: T CR LF
Accessible responses:
T_A CR LF
T_D CR LF
- command understood and in progress
- command carried out
T_A CR LF
T_v CR LF
- command understood and in progress
- command understood but tarring range exceeded
T_A CR LF
T_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable
measurement result
T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format: OT CR LF
Accessible responses: OT_TARA CR LF command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
O
T
space
tare
space
unit
space
CR
LF
Tare - 9 characters with right justification
Unit - 3 characters with left justification
Caution:
Tare value is always given in adjustment unit.
Set tare
Format: UT_TARE CR LF, where TAREtare value
Accessible responses:
UT_OK CR LF
- command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognized (tare format incorrect)
Caution:
Use dot in tare format as decimal point.