8. WEIGHING
Before start of measuring process or in case of essential change of ambient
conditions at a workstation (e.g. ambient temperature change at a workstation)
the moisture analyzer requires adjusting. The procedure of moisture analyzer
adjustment is described further in this user manual.
Before start of measuring procedure, it is recommended to load the
moisture analyzer’s pan a few times with mass close to the max
capacity,
Check if unloaded moisture analyzer indicates “precise zero” -
and whether measurement is stable -
, if not press ESC/TARE
key,
Place weighed object on moisture analyzer’s weighing pan and read
result only on stabilization of measurement result,
Mass indication of a load placed on moisture analyzer’s weighing pan
can be tarred for multiple times by pressing ESC/TARE key (pay
attention not to exceed maximal capacity of a moisture analyzer by
applying multiple tare function).
During times between carrying out the following measurement series do not
unplug the moisture analyzer from mains. It is recommended to switch off
moisture analyzer’s display by pressing ON/OFF key. On repeated pressing of
the ON/OFF key the moisture analyzer is ready for operation and carrying out
the following measurements.
9. MOISTURE ANALYZER MAIN PARAMETERS. ADJUSTING MOISTURE
ANALYZER OPERATION TO AMBIENT CONDITIONS AT A
WORKSTATION
Setting filtering level AuE.
Autozero function Auto.
Access to library of drying modes Libr.
Backlight of measurement result - bl.
Display brightness - blbA
“Beep” sound reaction on pressing a key bEEP.