17. USER CALIBRATION
Only for non-verified scales
Confirmation of high accuracy of weighing requires periodical correcting
of calibration factors in the scale memory – this is adjustment of the scale.
Calibration should be performed when we start weighing or dynamic
change of temperature occurs. Before starting calibration remove
loads from the pan.
17.1. Calibration
Procedure:
Enter submenu <P6.CAL> and then:
Following messages will be displayed:
During this time start mass is adjusted, and after completing the
procedure calibration weight mass is displayed (e.g. 3.000kg),
- 51 -