17.4. Troubleshooting
Problem: no reaction to pressing main switch on/off key (display remains dark).
Probable cause:
No voltage in mains
Damaged power cord
Damaged fuse of the moisture analyzer
Moisture analyzer defect
Problem: too long pending time for drying process finish.
Probable cause::
Selected finish mode of drying process is incorrect. Carry out experimental
tests to select a proper means of finish mode.
Problem: measurements are not repeatable
Probable cause:
non-uniform sample content. Prepare the sample from larger amount of
substance.
the drying time is too short. Change testing process finish mode criteria.
drying temperature is too high, causing sample’s oxidizing. Lower drying
temperature.
tested sample boils. Lower the drying temperature.
the temperature sensor is contaminated or defected. Clean the
temperature sensor.
the workbench on which the moisture analyzer is located is unstable.
Change the location of instrument’s operation.
- the ambient area is incompatible with the requirements (vibrations, drafts,
etc.). Change the ambient conditions for compatible ones according to the
guidelines of this user manual.