Radwag MYA 5.3Y.F1 Microbalance Uživatelský manuál Strana 110

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 198
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 109
- 110 -
<VALUE>, or the data on ambient conditions is automatically acquired from
the ambient conditions module and displayed for confirmation if option
<Ambient conditions> is set to <ONLINE>.
Caution:
If option <Ambient conditions> in the working mode is disabled set to
<No>, the software skips this part of the process.
On accepting the ambient conditions by pressing < > key, the software
returns to the mode’s main window and displays a command on unloading
the sample from the weighing pan (displayed in the message bar).
The user should unload weighed sample and the container from balance’s
weighing pan and accept the process by pressing < > key.
The software automatically moves to the following sample in a series.
Follow above procedure and weigh the following samples.
The process can be aborted by pressing < > key.
In case of a single series, and while weighing the following samples, the
measurements can be carried out according to the following outline: weighing
A, weighing T or weighing T+A. Each of samples saved in a series shall have a
description on its data, i.e. a status, which informs a user on current process
phase completed for a given sample. On entering the database of series and
previewing one of series, a window is opened with samples saves in the series.
Status denotation:
weighing A, measurement completed for procedure <Weighing A> or
<Weighing T+A>
weighing T, measurements completed for procedure <Weighing A>
Zobrazit stránku 109