Radwag MYA 5.3Y.F1 Microbalance Uživatelský manuál Strana 171

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 198
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 170
- 171 -
SUI_I CR LF
- command understood but not accessible at this
moment
MASS FORMAT
- mass value in current measuring unit is returned
immediately
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I stability space sign mass space unit CR LF
Example:
S U I CR LFcommand from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command carried out,
immediate response of mass value in current measuring unit
Where: _ - space
Switch on continuous transmission in basic measuring unit
Format: C1 CR LF
Accessible responses:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S I space
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in basic measuring unit
Format: C0 CR LF
Accessible responses:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and in progress
Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Accessible responses:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space character mass space unit CR LF
Zobrazit stránku 170
1 2 ... 166 167 168 169 170 171 172 173 174 175 176 ... 197 198

Komentáře k této Příručce

Žádné komentáře