Using <Minimum sample> mode guarantees that the results of completed
weighing processes are within set tolerance limits, and they comply with the
requirements of the quality system accepted by an organization.
Caution: the mode operates in the weighing mode only.
Available options:
METHOD
It is a means for designating an accepted quality standard.
Pressing a field displays a window with a list of minimal sample
weight values stored in balance’s memory, following a criterion on
determining the minimum sample.
Adding a new method is available only from the menu level
<Databases /Minimum samples>
MODE
Lock on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. Additionally the software disables accepting the measurement
result if it is below the minimum sample limit
Warn on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. The user can accept a measurement that is below the
minimum sample limit, but on a printout such measurement is
preceded by an asterisk (*).
Caution:
Means of setting a method of minimum sample determining is described
in point 27.9. of this user manual: <Databases /Minimum sample>.
RADWAG authorized personnel is only authorized to add new values of
minimum samples.
Weighing with application of <MINIMUM SAMPLE> mode
If during a weighing process a user wants to be informed whether a
measurement exceeds the minimum sample limit for a specific weighing
threshold, then the <Minimum sample> mode has to be enabled in the
additional setting of the weighing mode.
Procedure:
1. Press grey coloured workspace
2. The display indicates menu: Settings, Keys, Information, Printouts and
Profiles
3. Select menu <Settings> - additional options on weighing process
4. Press < Minimum sample> field
5. Below window is opened