- 134 -
4. Next control a field for entering expiry date of the minimum sample.
2 weeks before the expiry date the pictogram on minimum sample
status displayed in the mass measuring window section is
supplemented by a clock. It informs on the approaching expiry of the
minimum sample. Contact RADWAG representative to carry out
necessary changes in minimum sample settings.
5. Limits option that enables entering data on the mass value of a
minimum sample and range of packaging mass (tare) with applicable
criteria:
Taremaximum value of tare for which the minimum sample is valid.
the software enables entering three characteristic quantities: 0.000g,
optional mass from balance measuring range and Max balance
measuring range (see a below example).
Minimum mass the value of minimum mass that is determined for
a balance in its place of use and using appropriate methods.
Example no. 1 a balance AS 220.3Y, d=0.0001 g:
This balance has the following values of minimum sample determined for the
following values of packaging mass.
No.
Tare value
Minimum
sample
Means of operation
1
0.0000 g
0.1000 g
The minimum sample refers to all net
weighments that are weighed without a
packaging (the <TARE> button is not
applied)
2
10.0000 g
1.0000 g
The minimum sample refers to all net
weighments that are weighed in a
packaging that mass ranges from 0.0001 g
to 10.0000 g inclusive (the <TARE> button
applied)
3
50.0000 g
2.5000 g
The minimum sample refers to all net
weighments that are weighed in a
packaging that mass ranges from 10.0001
g to 50.0000 g inclusive (the <TARE>
button applied)
4
200.0000 g
4.0000 g
The minimum sample refers to all net
weighments that are weighed in a
packaging that mass ranges from 50.0001
g to 200.0000 g inclusive (the <TARE>
button applied)
Example no. 2 a balance AS 220.3Y:
No.
Tare value
Minimum
sample
Means of operation
1
220.0000 g
0.5000 g
The minimum sample refers to all net
weighments that are weighed in a
packaging with optional mass from full